UJI KESAMAAN BERKAS KOLIMATOR DENGAN VARIASI JARAK FFD (FOCUS FILM DISTANCE) PADA PESAWAT SINAR-X
Abstract
The collimator is a part of the x-ray device that has the function of regulating the size of the field. If this field is not aligned, the anatomy that should enter the irradiation field is not exposed to radiation. The large number of exposures performed can result in a mismatch between the area of the irradiation field and the X-ray beam produced. Distance variations are used to determine the suitability or value of the shift in the collimator field area tested. This study aims to determine the feasibility of a collimator on an X-ray plane by determining the similarity of the collimated beam with the X-ray beam.The type of research used is quantitative with experimental methods in the radiology installation of Andalas University Hospital Padang in August 2023 with variations of FFD 90 cm, 100 cm, 110 cm, and 120 cm. The measuring instrument used collimator test tool with collimation size 14x18 cm using small focal spot and large focal sp-axis (horizontal) and Y-axis (vertical) values by not exceeding 3% of FFD (ΔX + ΔY ≤ 3% of FFD).The test results of the collimator beam similarity test with FFD distance variations that comply with the tolerance limit ≤ 2% FFD are at FFD 100 cm, 110 cm, and 120 cm when using a small focal spot. While those that do not match or exceed the tolerance limit ≤ 2% FFD are FFD 90 cm small focal spot, FFD 90 cm, 100 cm, 110 cm, and 120 cm when using a large focal spot. Collimation that is feasible to use in the radiographic imaging process on a small focal spot with a distance (FFD) starting from 100 cm. The large focal spot is not feasible to use and needs to be improved.
Keywords: Conformity test, Beam collimation, FFD distance variation
Full Text:
PDFReferences
Akhadi, Muchlis. (2000). Dasar-dasar Proteksi Radiasi. Jakarta: Rineka Cipta.
Bushong, Stewart Carlyle. (2013). Radiologic Science for Technologists. Tenth Edition. Houston, Texas: Elsevier.
Carlton, R. R & Arlene, McKenna Adler. (2012). Principles of Radiographic Imaging: An Art and A Science. Delmar Cengage Learning.
Chadidjah S. 2012. “Penentuan ketepatan titik pusat berkas sinar Pada pesawat mobile x- ray sebagai parameter Kualitas kontrol di rsud. Prof. Dr. Hm. Anwar Makkatutu bantaeng”. Skripsi: FMIPA Universitas Hasanuddin Makassar.
Dendy, P. P & Heaton, B. (2012). Physics for Diagnostic Radiology. Boca Raton: Taylor & Francis Group.
Fauber, Terri L. (2013). Radiographic Imaging & Exposure. St. Louis: Elsevier.
Fransiska, E., dkk. (2018). “Uji Kesesuaian Berkas Sinar-X dengan Berkas Kolimator pada Pesawat
Sinar-X di Instalasi Radiologi RSUD Raden Mattaher Jambi”, dalam Jurnal Komunikasi Fisika Indonesia Program studi Fisika Fakultas Sains dan Teknologi Universitas Jambi, Vol. 15 No. 01 April 2018.
Keputusan Menteri Kesehatan Republik Indonesia Nomor 1250 Tahun 2009. Tentang Pedoman Kendali Mutu (Quality Control) Peralatan Diagnostik.
Papp, Jeffery. (2019). Quality Management in the Imaging Science.Sixth Edition. St Louis: Missouri.
Peraturan Kepala BAPETEN Nomor
Tahun 2014. Tentang Keselamatan Radiasi Dalam Produksi Pesawat Sinar-X Radiologi Diagnostik dan Intervensional.
Rahman, Nova. (2009). Radiofotografi. Padang: Universitas Baiturrahmah.
Sari, Ayu Wita & Siti Hartina. (2017). “Uji Kesesuaian Collimator Beam Dengan Berkas Sinar-X pada Pesawat Raico di Instalasi Radiologi Raden Mattaher Jambi”, dalam Prosiding Pertemuan dan Presentasi Ilmiah Peneltian.
Savitri, Retno Indah, dkk. (2014). “Optimasi Faktor Eksposi pada Sistem Radio Optimasi Faktor
Eksposi pada Sistem Radiografi Digital Menggunakan Analisis CNR (Contrast to
Noise Ratio)”, dalam Unnes Physics Journal, Vol. 3 No. 1 2014.
DOI: http://dx.doi.org/10.30633/jsm.v7i2.2849
Refbacks
- There are currently no refbacks.
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Jurnal Kesehatan Saintika Meditory
Published by Stikes Syedza Saintika Padang
Pusat Penelitian dan Pengabdian Masyarakat, STIKes SYEDZA Saintika Padang
Jl. Prof. Dr. Hamka No. 228 Air Tawar Timur Padang - Sumatera Barat
Phone: 082384992512
Email: lppmsyedza@gmail.com
View My Stats